Pacemakers and ICDs in the operating room

Cardiovascular and cardiac anesthesia·~4 min read· Physician-reviewed✓ Sourced · 2

Stem

A 74-year-old man, 78 kg, is scheduled for open sigmoid colectomy through a midline incision. He has ischemic cardiomyopathy with an ejection fraction of 30% and complete heart block, and has a biventricular ICD in the left chest. His ECG shows 100% paced complexes at 70. The device was last checked 9 months ago, and he has had no shocks.

Question

How will you manage his device before, during and after surgery?

Ideal answer

HeadlinePacemaker-dependent with an ICD: monopolar cautery above the umbilicus can inhibit pacing into asystole or trigger an inappropriate shock.1 I get the device team in.

  1. Identify the device and settings; interrogate; get a prescription, because an ICD should be checked within 6 months and his is 9 months out.1
  2. Reprogram to asynchronous pacing, tachytherapy suspended, at a rate that covers his needs; a magnet on an ICD turns off shocks but has no effect on pacing.1
  3. External defibrillation and pacing pads before induction, with a plethysmograph or arterial line so I see perfusion when cautery obscures the ECG.
  4. Surgeon: dispersive pad placed so current avoids the generator, bipolar or ultrasonic instruments where possible, short bursts.2
  5. Continuous monitoring with pads until reprogrammed and interrogated, restoring his original settings after surgery.2

Breaking down the answer

  • Dependence decides the plan — if EMI inhibits pacing in a dependent patient, the result is severe bradycardia or asystole.2 An underlying rhythm on a pacing check or a history of heart block answers the question.
  • Above versus below the umbilicus — below, with a correctly placed pad, there is minimal need to reprogram or use a magnet; above, clinically important EMI occurred in 20% of noncardiac cases in one study.1
  • Magnet behavior is device-specific — on most pacemakers a magnet gives asynchronous pacing at a set rate that may not meet his needs, and must be confirmed for the device.2 The 2020 ASA advisory discourages indiscriminate magnet use.1
  • Distance — cautery within about 6 inches of the generator is the highest risk for inhibition or inappropriate therapy.1
  • Rate-responsive features off — minute-ventilation sensors misread ventilation and EMI as exercise and pace fast; the device team disables them.
  • Re-interrogation triggers — cardioversion, defibrillation, large blood loss or electrolyte shifts.2 The plan has an end point, not just a start.

Follow-up question

Two hours into the case, with the ICD therapies suspended, the monitor shows ventricular fibrillation and the arterial trace goes flat.

What do you do?

Follow-up answer

HeadlineVentricular fibrillation the suspended ICD cannot treat. I call it out, stop cautery, call for help and the code cart.

The pads are already on, so I defibrillate immediately with the external defibrillator and run CPR and ACLS between shocks. Had a magnet suspended therapy, I would remove it and watch briefly for the ICD to treat, without delaying an external shock.1 After return of circulation, the cause: ischemia, potassium, magnesium, acidosis. I tell the device team, because the device needs interrogation after an external shock, and he stays monitored with pads until his settings are restored.2

Pitfalls that cost points

  • Putting a magnet on an ICD and assuming the patient is now safely paced.
  • Leaving recovery with tachytherapy still off.

Rapid-fire

Q: How often should a pacemaker and an ICD be checked? A: Pacemakers yearly; ICDs and CRT devices every 6 months.1

Q: How should a leadless pacemaker be managed when EMI is expected? A: Reprogram it to VOO asynchronous pacing.1

Q: How do you confirm a magnet is working on a subcutaneous ICD? A: Listen for the audible beep; if absent, reposition it or reprogram.1

Q: What does the fourth letter of the pacemaker code describe? A: Rate modulation, the ability to change rate to meet physiologic demand.2

Q: How much battery life should the device have? A: At least 3 months.2

Numbers to know

Item Value
Pacemaker interrogation Yearly1
ICD or CRT interrogation Every 6 months1
EMI above umbilicus Clinically important in 20%1
High-risk cautery distance Within 6 inches of generator1
Minimum battery life 3 months2

Sources

  1. 1.Change of Pace: An Update on the Perioperative Management of Cardiovascular Implantable Electronic Devices (CIEDs) · APSF · apsf.orgSummarizes the 2011 HRS/ASA expert consensus and the 2020 ASA practice advisory (Anesthesiology 2020;132:225-252), whose full text was not reachable from the fetcher.
  2. 2.Cardiac implantable electronic devices: pacemakers · OpenAnesthesia · openanesthesia.orgStates the ASA advisory recommends interrogation 3-6 months before surgery; the APSF summary of the HRS/ASA consensus gives yearly for pacemakers and 6 months for ICD/CRT. The page uses the HRS/ASA intervals.

Facts on this page were checked against these sources on Oct 15, 2026. Guidelines change; confirm doses against current institutional protocols.

Found an error or a better way your examiners phrased this? Send feedback on this page.